Take Home Lesson - I

Knowledge of correct lineshapes allows you to decide quickly whether your sample is correctly shimmed. You have to decide whether the return (a better lineshape) is worth the time spent achieving that lineshape.

 

image015

image017

III. Shimming, Linewidths and Lineshapes -III

Knowledge of correct lineshape can help in correcting problems such as those shown in Figure III-3. Although the peak in Figure III-3b may have a linewidth at half-height (LW1/2) that is less than 0.50 Hz, the checkout requirement, it is obviously poorly shimmed. You should never accept a poorly shimmed lineshape as is shown in Figure III-3b, where a single line is expected.

On the pages that follow are some lineshape defects and the shims that should be adjusted to correct the problem. In general, odd-order shims (Z1, Z3) affect the lineshape symmetrically while eve-order shims (Z2, Z4) cause a non-symmetrical lineshape. The higher the order (Z4 is higher order than Z2), the lower (i.e., closer to the base of the peaks) the problem is observed.

Reblog this post [with Zemanta]

III. Shimming, Linewidths and Lineshapes - II

It is important to have a basic understanding of lineshape to be able to judge when:

(1) The shimming is off, and

(2) More time should be spent in shimming the sample. The best way to avoid problems is to establish a procedure using the guidelines below.

1. Always load a shim file when you sit down at the instrument. You should never assume the previous user left the instrument with a standard shim file loaded. Without reloading standard shims, you will have to start where the last person stopped --- and that might include someone who shimmed for a short sample, a “bad” tube, a viscous sample, etc.

2. Be aware of lock parameters, especially if you only shim on the lock display. Establish lock transmitter power and gain levels that work for most of your samples. If you encounter a sample that seems to require an unusually high power or gain setting, there is a problem with your sample and/or the instrument, and shimming on the lock level may be difficult or impossible.

3. Shimming problems are confirmed only if the “problem” is visible on every peak in your spectrum. If only one peak is doubled, the “problem” is sample related, and can’t be shimmed away. Remember, anomalies close to the base of intense single lines may not be visible on less intense peaks unless the vertical scale is increased.

4. Establish a shimming method. Shimming is an art form that requires patience and practice. You should always approach shimming with some method that works for you to give acceptable results. Example: load a shim file, adjust lock level to a maximum with Z1, then Z2, then Z1, then Z3, then Z1.

5. Spinning sidebands should always be below 2%. If spinning sidebands are above 2%, turn off the spinner air, optimize X and Y, then turn the spinner air back on and re-optimize Z1, Z2, and Z3. If this does not solve the problem, consider transferring your sample to another tube.

Reblog this post [with Zemanta]

III. Shimming, linewidths, and lineshapes - I

It is important to have a knowledge of proper lineshape of an NMR signal and how improper shimming contributes to poor lineshape. Examples of reasonably well-shimmed NMR spectra from the Unity+300 and VXR-S NMR spectrometers are illustrated in Figure III-1 and III-2.

NMR peaks have a Lorenzian lineshape. The mathematical expression of a Lorentzian line and the three parameters associated with the line are shown below.

clip_image002

The minimum obtainable linewidth at half-height is directly related to the resolution of an instrument, i.e., how close two peaks can be and still be distinguishable. Resolution is usually measured using o-dichlorobenzene, which has very narrow lines its 1H NMR spectrum. The manufacturer’s resolution specification is usually around 0.20 Hz, although linewidths of less than 0.10 Hz are obtainable with expert shimming.

Manufacturers of NMR instruments, however, have traditionally separated the resolution specification from the lineshape specification. Lineshapes for 1H NMR spectra are usually specified using CHCl3 and the specifications are stated in terms of linewidth at 0.55% and 0.11% height of the CHCl3. These percentages are chosen because they are the height of the 13C satellites of the CHCl3 line and one-fifth this height. However, these values are meaningful only when compared with the half-height width (LW1/2). From the mathematical equation for a Lorentzian line, the linewidth at 0.55% height is calculated to be 13.5 times the LW1/2, while the linewidth at 0.11% height is calculated to be 30 times the LW1/2. So, if the linewidth at half-height is 0.30 Hz, the calculated values are 4.0 Hz at 0.55% and 9.0 Hz at 0.11%.

clip_image004

For comparison, the manufacturer’s specifications are 10-15 Hz and 20-30 Hz at 0.55% height and 0.11% height respectively. These values are larger than the theoretical values because the linewidths at 0.55% and 0.11% height are very sensitive to how well shimmed the magnet is.

Reblog this post [with Zemanta]