IV. Signal-to-Noise Measurement - II

Choice of a noise region must be consistently applied for standard samples, and for 0.1% ethylbenzene (ETB), we have chosen 5 to 3.5 ppm. It is also critical that the noise be amplified if necessary in order to make an accurate measurement using a ruler. There is not much point in

trying to compare the ratio of noise to a few millimeters high to a peak 20 cm high; plot the noise such that it is several centimeters high. If the instrument performs the signal-to-noise measurement and calculation, the relative intensities are not as critical. Both the Unity+300 and VXR-S 400 NMR spectrometers perform the S/N calculation by software commands.

The last point concerning sensitivity is demonstrated during the 13C{1H} part of the checkout. The checkout requires that you acquire three spectra, using 1, 4, and 16 transients. Signal-to-noise increases as the square root of the number of transients; to double the signal-to-noise you must take four times as many transients.

When using a concentrated sample such as 30% menthol for 13C, or when running routine 1H spectra, the number of transients or scans is quite small, so the point discussed above may not seem important. However, suppose you are in the following situation: you have only a few mg of research sample, and after collecting a 13C{1H} spectrum on the Unity+300 for 2 hours, you get peaks with S/N of only 5:1. Since the peaks are barely visible above the noise ( and you may have missed quarternary carbons), you want to re-collect the spectrum to get a S/N of 50:1, a value most typical for carbon NMR. Unfortunately, this will take 10x10x2 = 200 hours! In such cases, you should consider using a higher field strength spectrometer, which will be able to give you a spectrum with good S/N in much less time.

 

Reblog this post [with Zemanta]

IV. Signal-to-Noise Measurement - I

               The first spectrum described in the checkout involves determining a signal-to-noise measurement, or S/N. S/N is an important criterion for accurate integrations (see section V), and is also one of the best ways to determine how sensitive an NMR spectrometer is. In general, a higher S/N specification means that the instrument is more sensitive.

A typical result for the Unity+300 checkout is S/N = 87 and is shown in Figure IV-1, while a typical result for he VXR-S 400 checkout is S/N = 109 and is shown in Figure IV-2. If you measure S/N for the ethylbenzene standard that is significantly lower than these values, write your observations in the spectrometer logbook and inform S. Chandrasekaran .

S/N measurements for proton spectra are always determined using a sample of 0.1% ethylbenzene in CDCl3 (ETB). It is important that the spectrum be acquired under standard conditions:

1. 900 pulse width

2. Line Broadening = 1.0 Hz

3. Spectral Width = 15 to 5 ppm

4. A sufficient relaxation delay (at least 5xT1)

5. A sufficient digital resolution (less than 0.5 Hz/point)

6. One transient acquisition

Optimum signal-to-noise ratio for any sample is achieved using a line broadening equal to the linewidth at one-half height. When this line broadening is applied, the linewidth at half-height doubles, i.e., it is the sum of the natural linewidth at one-half height plus the line broadening applied. The equation used for calculating S/N is:

S/N = (2.5 x A) / Npp

A = height of the chosen peak Npp = peak-to-peak noise

Peak-to-peak noise means exactly that - a measurement from the most positive to the most negative positions for the noise.

As shown below, the wildest differences occur with low probability, so there seems to be a large gap between them and what appears to be the main part of the noise; will-power must be exercised in order to truly measure this gap and not to dismiss it in our minds as “spikes” rather than the incoherent noise it is. True instrumental defects resulting in spikes can be dismissed, but only after several measurements have proven a spike is always present.