Components: Relaxation Delay [d1]

While running repetitive transients (nt > 1, e.g.,) to obtain high signal-to-noise ratio in an NMR experiment, the time between each transient becomes important. This time, called the pulse repetition time (prt), also called recycle time should be long enough to allow complete relaxation of the nuclear spins between transients. Incomplete relaxation will lead to loss of signal and is one cause of inaccurate integrations. The pulse repetition time is given by the following equations.

prt = Pulse Width + Acquisition Time + Relaxation Delay or

prt = pw + at + d1

The relaxation time for a nucleus is called T1 (longitudinal relaxation time) and if a 90 degree pulse is used to excite the spins (Figure II-2A), a pulse repetition time equal to or greater than 5xT1 is required to have complete relaxation (see Figure II-5). If a pulse width less than 90 degree pulse width is used, the pulse repetition can be proportionally less. In the above equation, pw is in

clip_image002

microseconds whereas at and d1 are in seconds. Thus, pw can be ignored in the calculation of the pulse repetition time. In practice, on the Unity+300 and VXR-S NMR spectrometers, at will be a fixed parameter chosen for the resolution desired, and therefore, the user has the only option of setting the relaxation delay d1 to a proper value so that the pulse repetition time can be optimum.

For example, if T1 is 0.5 sec, the optimum prt should be 2.5 sec for a 90 degree pulse used for excitation. The standard parameters for the Unity+300 and VXR-S 400 contain at set to 4.096 sec and 4.099 sec. Since these values are greater than 5xT1, d1 could be set to 0 to meet the condition that prt be equal to or greater than 5xT1. The standard parameters on the Unity+300 and VXR-S 400 NMR spectrometers for 1H and 13C include 45 degree pulse width for excitation and the proper d1 to give a prt of 5.1 sec which will be optimum for most routine NMR runs, but exceptions are possible!

Reblog this post [with Zemanta]

Components: Number of Points [np] and Spectral Width [sw]

Performing a mathematical operation called the Fourier transformation (ft) on the FID (time-domain data) generates a spectrum which gives intensity of signals as a function of frequency (frequency-domain data), as shown in Figure II-3. This familiar NMR spectrum is characterized by

clip_image002

two important parameters, the spectrometer frequency (sf), and the spectral width (sw). As

shown in Figure II-4, at a spectrometer frequency of 299.957 Hz on the Unity+300, a spectral width (sw) of approximately 3000 Hz is required to observe 10 ppm (10 ppm x 300 Hz/ppm). At a spectrometer frequency of 399.964 MHz on the VXR-S 400, a spectral width (sw) of

approximately 4000 Hz is required to observe the same 10 ppm (10ppm x 400 Hz/ppm)

clip_image004

The spectral width sw in Hz, acquisition time at in seconds and the number of points np are related to each other by the following equation:

at = np/2sw equation 1

The digital resolution is related to acquisition time at as follows.

resolution = 1/at = 2sw/np equation 2

From equation 2, it follows that resolution or digital resolution is in the units of Hz/point. To obtain properly digitized NMR spectra, the digital resolution must be less than one-half the peak width at half-height of the narrowest peak in the NMR spectrum. Using this optimum digitization condition guarantees that each peak is represented by at least 3 points. For example, if the narrowest peak width is 0.7 Hz at half-height, the required digital resolution is less than 0.35 Hz. It follows that for a spectrometer frequency of 399.964 MHz, a spectral width of 4000Hz (10 ppm) and a digital resolution of 0.35 Hz/point, the number of points can be calculated from equation 2 as follows:

np = 2sw/resolution ; np = 2*4000/0.35 = 22857 points

Normally, the spectral width is fixed for a given NMR experiment and either the number of points or the acquisition time is modified to achieve the desired digital resolution. On both the Unity+300 and VXR-S 400 NMR spectrometers, it is best to modify acquisition time by setting at to the proper value to achieve the resolution (see equation 2). The spectrometer will automatically set the correct number of points np. The standard parameters for 1H observation are as follows.

Unity+300 VXR-S 400

Spectral Width (sw) = 4000 Hz (13 ppm) Spectral Width (sw) = 5199.5 Hz (13 ppm)

Acquisition Time (at) = 4.096 sec Acquisition Time (at) = 4.099 sec

Number of points (np) = 32768 Number of Points = 42624

Digital Resolution = 0.24 Hz/point Digital Resolution = 0.24 Hz/point

Reblog this post [with Zemanta]

Components: Acquisition Time [at]

After the radiofrequency pulse has been applied to the sample and turned off, the spins return to thermal equilibrium, and in the process induce a signal in the receiver coil. In Figure II-1, the decaying sine wave depicts this signal called the Free Induction Decay or FID. The FID is a plot of the intensity of the induced signal as a function of time, and is commonly referred to as the time-domain data. The time spent in acquiring the FID is called the acquisition time (at). The FID, which is an analog signal is digitized using an analog-to-digital converter which represents the FID by a series of points along the FID curve. The number of points thus used can be entered by setting np to the proper value on the Unity+300 and VXR-S 400 NMR spectrometers. As a general rule, more points result in higher resolution.

Reblog this post [with Zemanta]

Components: Pulse Width [pw]

If the average equlibrium magnetization of all the spins in a sample placed in a magnetic field is represented as oriented along the Z axis in a cartesian coordinate system as shown in Figure II-2, then the length of time the radiofrequency pulse (rf energy) is applied along the X axis (for

clip_image002

example) will have a specific effect on the magnetization. The length of time the radiofrequency pulse is applied to the sample is called the pulse width (pw). The pulse width is normally defined in terms of a flip angle in degrees. As shown in Figure II-2, different pulse widths have different effects on the equilibrium magnetization. The 90 degree pulse width is the length of time the

pulse of rf energy is applied to a particular sample to flip (rotate) all the spins from the Z axis into the XY plane, as shown in Figure II-2A. The 90 degree pulse width, specified as pw90 on the Unity+300 and the VXR-S 400 NMR spectrometers depends upon the nucleus, solvent, and the instrument (probe, etc.,). On the Unity+300, pw90 is 13.3 microseconds for proton and 9.7 microseconds for carbon. On the VXR-S 400, pw90 is 35.0 microseconds for proton and 13.0 microseconds for carbon. Routine experiments on these spectrometers are performed using a pulse width much less than the 90 degree pulse width. The pulse width is entered in microseconds by setting pw to the proper value on both the NMR spectrometers.

Reblog this post [with Zemanta]

Components: Spectrometer Frequency [sf]

The nuclei are excited by the radiofrequency pulse from a transmitter and upon returning back to thermal equilibrium during acquisition time induce a signal in the receiver coil which is detected as the NMR signal. The radiofrequency pulse has a characteristic frequency, named the spectrometer frequency (sf), which is dependent on the nucleus observed and the magnetic field strength of the NMR spectrometer used. For protons, sf on the Unity+300 is approximately 299.957 MHz, and on the VXR-S 400, it is 399.964 MHz. For carbons, sf on the Unity+300 is approximately 75.432 MHz, and on the VXR-S 400, it is 100.580 MHz. The spectrometer frequency specifies the center of the NMR spectrum acquired.

Reblog this post [with Zemanta]

Components of the Pulse FT-NMR Experiment

This section will cover the most important parameters which affect any spectrum you may collect using an FT-NMR spectrometer. The standard names of the parameters and the specific names (in bold) by which these parameters are referred to on the Unity+300 and VXR-S 400 NMR spectrometers are listed below.

Spectrometer Frequency sf

Pulse Width pw

Acquisition Time at

Number of Points np

Spectral width sw

Relaxation Delay d1

clip_image002

The generic 1PULSE FTNMR experiment is illustrated in Figure II-1. The experiment essentially

Consists of three components: relaxation delay, pulse width and acquisition time, and is named a 1PULSE experiment because one radiofrequency (rf) pulse is applied per cycle or transient.

Reblog this post [with Zemanta]